UNIQUE MEASUREMENT PROCEDURE WITH COMBICUFF.
In the event that a patients limb circumference is borderline between |
one bladder size and another ![]() |
measurement with first one bladder size and then the other without |
unduely disturbing the patient. |
1. Firstly, activate the smaller bladder by pressing down the smaller |
size on the SELECTION VALVE then measure the patients blood |
pressure and record the result. |
2. Secondly, activate the larger bladder by pressing down the larger size |
on the selection valve and repeat and record the measurement. If it is |
deemed necessary an average of the two results can be made. |
Helsingborgsv. 33 | Tel: +46 431 - 833 15 |
262 72 Engelholm | Fax: +46 431 - 833 33 |
Sweden | Email: pdc@polymer-dip.com |
Copyright © PDC AB
pdc@polymer-dip.com